Recall Z-1170-2018— St. Jude Medical, Inc.

Class II · Terminated

Class II FDA medical device recall by St. Jude Medical, Inc., initiated 2018-03-09, terminated 2020-11-03. FDA's recall record names no specific 510(k) clearance for this event. Product: Proclaim DRG Implantable Pulse Generator, Model Number 3664. Reason: The firm received complaints of error messages that occurred during routine impedance checks on Proclaim DRG IPGs, model 3664. Some complaints were also associated with transient over stimulation which created discomfort for the patients..

Recalling firm
St. Jude Medical, Inc.
Classification
Class II
Status
Terminated
Initiated
2018-03-09
Terminated
2020-11-03
Firm location
Plano, TX, United States

Product description

Proclaim DRG Implantable Pulse Generator, Model Number 3664

Reason for recall

The firm received complaints of error messages that occurred during routine impedance checks on Proclaim DRG IPGs, model 3664. Some complaints were also associated with transient over stimulation which created discomfort for the patients.