Recall Z-1166-2023— Siemens Medical Solutions USA, Inc
Class II · Ongoing
Class II FDA medical device recall by Siemens Medical Solutions USA, Inc, initiated 2022-12-21. It names one FDA 510(k) clearance: K062623. Product: Luminos dRF with software VD10- A diagnostic imaging system for radiographic and fluoroscopic studies Model Number: 10094200. Reason: Potential risk of collision with the ceiling, wall, or objects which may result in serious injury to staff or patients due to room configuration parameters of the system being set to default values.
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2022-12-21
- Firm location
- Malvern, PA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Luminos dRF with software VD10- A diagnostic imaging system for radiographic and fluoroscopic studies Model Number: 10094200
Reason for recall
Potential risk of collision with the ceiling, wall, or objects which may result in serious injury to staff or patients due to room configuration parameters of the system being set to default values
Data sourced from openFDA. This site is unofficial and independent of the FDA.