Recall Z-1164-2025— Microbiologics Inc

Class II · Ongoing

Class II FDA medical device recall by Microbiologics Inc, initiated 2025-01-20. FDA's recall record names no specific 510(k) clearance for this event. Product: Microbiologics Multi-Strain Human Papillomavirus (HPV) Verification Panel (Inactivated Pellet), REF 8235; in-vitro diagnostic test. Reason: Internal quality control procedures were not followed correctly for the testing/release of three lots of 8234/8235. This resulted in the release of products that should have failed and been discarded. The concentration of one target (HPV 16 L1) was low enough to cause QC failures, no detection, or invalid results on some instruments and assays..

Recalling firm
Microbiologics Inc
Classification
Class II
Status
Ongoing
Initiated
2025-01-20
Firm location
Saint Cloud, MN, United States

Product description

Microbiologics Multi-Strain Human Papillomavirus (HPV) Verification Panel (Inactivated Pellet), REF 8235; in-vitro diagnostic test

Reason for recall

Internal quality control procedures were not followed correctly for the testing/release of three lots of 8234/8235. This resulted in the release of products that should have failed and been discarded. The concentration of one target (HPV 16 L1) was low enough to cause QC failures, no detection, or invalid results on some instruments and assays.

Data sourced from openFDA. This site is unofficial and independent of the FDA.