Recall Z-1163-2020— Pacific Medical Group Inc.
Class II · Ongoing
Class II FDA medical device recall by Pacific Medical Group Inc., initiated 2019-09-11. It names one FDA 510(k) clearance: K012718. Product: External Toco Transducer (M1355A), that may have been serviced with the following: Pacific Medical Philips M1355A Toco Bottom Case, NFPH9315; Pacific Medical Philips M1355A Toco Cable Assembly, NFPH9320; Pacific Medical Philips M1355A Toco Gauge, NFPH9340. Reason: Fetal/maternal monitor/monitoring systems serviced or remanufactured using non-OEM equivalent components that have not been appropriately verified or validated. Possible adverse clinical impact: inability to accurately detect and measure fetal heart rate, inability to accurately detect maternal peripheral oxygen saturation, inability to monitor uterine activity to follow contractions, inability to determine temperature, and maternal blood pressure, lack of electrocardiography tracing, patient/clinician burns, and delay in detecting maternal or fetal distress..
- Recalling firm
- Pacific Medical Group Inc.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2019-09-11
- Firm location
- San Clemente, CA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
External Toco Transducer (M1355A), that may have been serviced with the following: Pacific Medical Philips M1355A Toco Bottom Case, NFPH9315; Pacific Medical Philips M1355A Toco Cable Assembly, NFPH9320; Pacific Medical Philips M1355A Toco Gauge, NFPH9340
Reason for recall
Fetal/maternal monitor/monitoring systems serviced or remanufactured using non-OEM equivalent components that have not been appropriately verified or validated. Possible adverse clinical impact: inability to accurately detect and measure fetal heart rate, inability to accurately detect maternal peripheral oxygen saturation, inability to monitor uterine activity to follow contractions, inability to determine temperature, and maternal blood pressure, lack of electrocardiography tracing, patient/clinician burns, and delay in detecting maternal or fetal distress.
Data sourced from openFDA. This site is unofficial and independent of the FDA.