Recall Z-1161-2016— Zimmer Manufacturing B.V.
Class II · Terminated
Class II FDA medical device recall by Zimmer Manufacturing B.V., initiated 2016-01-11, terminated 2017-04-13. It names one FDA 510(k) clearance: K934765. Product: Product 63 consists of all product under product code: LPH and same usage: Item no: 620004020 TRILOGY ACET SHELL 40MM O 620004220 TRILOGY ACET SHELL 42MM O 620004420 TRILOGY ACET SHELL 44MM O 620004620 TRILOGY ACET SHELL 46MM O 620004621 TRILOGY ACET SHELL 46MM O 620004820 TRILOGY ACET SHELL 48MM O 620004822 TRILOGY ACET SHELL 48MM O 620004823 TRILOGY ACET SHELL 48MM O 620005020 TRILOGY ACET SHELL 50MM O 620005021 TRILOGY ACET SHELL 50MM O 620005022 TRILOGY ACET SHELL 50MM O 620005220 TRILOGY ACET SHELL 52MM O 620005221 TRILOGY ACET SHELL 52MM O 620005222 TRILOGY ACET SHELL 52MM O 620005223 TRILOGY ACET SHELL 52MM O 620005420 TRILOGY ACET SHELL 54MM O 620005421 TRILOGY ACET SHELL 54MM O 620005422 TRILOGY ACET SHELL 54MM O 620005423 TRILOGY ACET SHELL 54MM O 620005620 TRILOGY ACET SHELL 56MM O 620005621 TRILOGY ACET SHELL 56MM O 620005622 TRILOGY ACET SHELL 56MM O 620005623 TRILOGY ACET SHELL 56MM O 620005820 TRILOGY ACET SHELL 58MM O 620005821 TRILOGY ACET SHELL 58MM O 620005822 TRILOGY ACET SHELL 58MM O 620005823 TRILOGY ACET SHELL 58MM O 620006020 TRILOGY ACET SHELL 60MM O 620006022 TRILOGY ACET SHELL 60MM O 620006023 TRILOGY ACET SHELL 60MM O 620006021 TRILOGY ACET SHELL 60MM OD SOLID 620006220 TRILOGY ACET SHELL 62MM O 620006221 TRILOGY ACET SHELL 62MM O 620006222 TRILOGY ACET SHELL 62MM O 620006420 TRILOGY ACET SHELL 64MM O 620006422 TRILOGY ACET SHELL 64MM O 620006620 TRILOGY ACET SHELL 66MM O For use in total hip arthroplasty. Reason: As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile..
- Recalling firm
- Zimmer Manufacturing B.V.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2016-01-11
- Terminated
- 2017-04-13
- Firm location
- Mercedita, PR, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Product 63 consists of all product under product code: LPH and same usage: Item no: 620004020 TRILOGY ACET SHELL 40MM O 620004220 TRILOGY ACET SHELL 42MM O 620004420 TRILOGY ACET SHELL 44MM O 620004620 TRILOGY ACET SHELL 46MM O 620004621 TRILOGY ACET SHELL 46MM O 620004820 TRILOGY ACET SHELL 48MM O 620004822 TRILOGY ACET SHELL 48MM O 620004823 TRILOGY ACET SHELL 48MM O 620005020 TRILOGY ACET SHELL 50MM O 620005021 TRILOGY ACET SHELL 50MM O 620005022 TRILOGY ACET SHELL 50MM O 620005220 TRILOGY ACET SHELL 52MM O 620005221 TRILOGY ACET SHELL 52MM O 620005222 TRILOGY ACET SHELL 52MM O 620005223 TRILOGY ACET SHELL 52MM O 620005420 TRILOGY ACET SHELL 54MM O 620005421 TRILOGY ACET SHELL 54MM O 620005422 TRILOGY ACET SHELL 54MM O 620005423 TRILOGY ACET SHELL 54MM O 620005620 TRILOGY ACET SHELL 56MM O 620005621 TRILOGY ACET SHELL 56MM O 620005622 TRILOGY ACET SHELL 56MM O 620005623 TRILOGY ACET SHELL 56MM O 620005820 TRILOGY ACET SHELL 58MM O 620005821 TRILOGY ACET SHELL 58MM O 620005822 TRILOGY ACET SHELL 58MM O 620005823 TRILOGY ACET SHELL 58MM O 620006020 TRILOGY ACET SHELL 60MM O 620006022 TRILOGY ACET SHELL 60MM O 620006023 TRILOGY ACET SHELL 60MM O 620006021 TRILOGY ACET SHELL 60MM OD SOLID 620006220 TRILOGY ACET SHELL 62MM O 620006221 TRILOGY ACET SHELL 62MM O 620006222 TRILOGY ACET SHELL 62MM O 620006420 TRILOGY ACET SHELL 64MM O 620006422 TRILOGY ACET SHELL 64MM O 620006620 TRILOGY ACET SHELL 66MM O For use in total hip arthroplasty
Reason for recall
As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
Data sourced from openFDA. This site is unofficial and independent of the FDA.