Recall Z-1155-2017— Biomet U.K., Ltd.

Class II · Terminated

Class II FDA medical device recall by Biomet U.K., Ltd., initiated 2016-01-05, terminated 2018-06-30. FDA's recall record names no specific 510(k) clearance for this event. Product: Oxford Partial Knee System Phase 3 Tibial Template Right Medial Size A Product Usage: Instruments for use with the Oxford Uni Partial Knee System. Reason: Zimmer Biomet is conducting a medical device field action for various lot-specific Oxford Knee instruments due to potential alumina inclusions in the raw material batch used to produce the affected products. Inclusions contained within the finished product could lead to the cracking and separation of the instrument..

Recalling firm
Biomet U.K., Ltd.
Classification
Class II
Status
Terminated
Initiated
2016-01-05
Terminated
2018-06-30
Firm location
Bridgend South Wales, N/A, United Kingdom

Product description

Oxford Partial Knee System Phase 3 Tibial Template Right Medial Size A Product Usage: Instruments for use with the Oxford Uni Partial Knee System

Reason for recall

Zimmer Biomet is conducting a medical device field action for various lot-specific Oxford Knee instruments due to potential alumina inclusions in the raw material batch used to produce the affected products. Inclusions contained within the finished product could lead to the cracking and separation of the instrument.

Data sourced from openFDA. This site is unofficial and independent of the FDA.