Recall Z-1151-2021— Baxter Healthcare Corporation
Class II · Terminated
Class II FDA medical device recall by Baxter Healthcare Corporation, initiated 2021-01-07, terminated 2024-04-30. It names one FDA 510(k) clearance: K190910. Product: The PrisMax System Version II Hemodialysis Delivery System: Automatic Repositioning System Pump Assembly, specifically tubing, Model Numbers: a) 955626 b) 955558 c) 955701 PrisMax V2 Preventive Maintenance Kit, Product Code SC616 ARPS PUMP, Product Code SC6035 ARPS Pump Assembly, Product Code SC6053. Reason: Variability in the performance of the tubing in the ARPS (Automatic Repositioning System) Pump Assembly which may lead to alarm situations during system self-test and therapy. This tubing is normally replaced during annual preventative maintenance. Alarm Situations: During System Self-Test: Alarm code B1215 (ARPS Self-Test failure); During Therapy: B1273 (ARPS Leak), B1215 (ARPS Pressure self-test), B1116 (Return Pressure self-test Failure), B1539 (Access Pod Reposition Failure), B1541 (Effluent Pod Reposition Failure), B1540 (Effluent Filter Pod Reposition Failure). If an alarm occurs, it may lead to delay or interruption of therapy. In the event that therapy is terminated without returning blood to the patient, blood loss may occur..
- Recalling firm
- Baxter Healthcare Corporation
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2021-01-07
- Terminated
- 2024-04-30
- Firm location
- Deerfield, IL, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
The PrisMax System Version II Hemodialysis Delivery System: Automatic Repositioning System Pump Assembly, specifically tubing, Model Numbers: a) 955626 b) 955558 c) 955701 PrisMax V2 Preventive Maintenance Kit, Product Code SC616 ARPS PUMP, Product Code SC6035 ARPS Pump Assembly, Product Code SC6053
Reason for recall
Variability in the performance of the tubing in the ARPS (Automatic Repositioning System) Pump Assembly which may lead to alarm situations during system self-test and therapy. This tubing is normally replaced during annual preventative maintenance. Alarm Situations: During System Self-Test: Alarm code B1215 (ARPS Self-Test failure); During Therapy: B1273 (ARPS Leak), B1215 (ARPS Pressure self-test), B1116 (Return Pressure self-test Failure), B1539 (Access Pod Reposition Failure), B1541 (Effluent Pod Reposition Failure), B1540 (Effluent Filter Pod Reposition Failure). If an alarm occurs, it may lead to delay or interruption of therapy. In the event that therapy is terminated without returning blood to the patient, blood loss may occur.
Data sourced from openFDA. This site is unofficial and independent of the FDA.