Recall Z-1151-2018— Beckman Coulter Inc.

Class II · Terminated

Class II FDA medical device recall by Beckman Coulter Inc., initiated 2018-01-18, terminated 2023-09-27. FDA's recall record names no specific 510(k) clearance for this event. Product: AutoMate 2550 "High Speed" Catalog Number ODL25255; AutoMate 1250 "Standard" Catalog Number ODL25125. Chemistry analyzer sample processing and sorting system.. Reason: A misaligned pipette tip sensor poses the remote risk of cross-contamination when failing to cause a no-tip error, resulting in pipetting patient samples without pipette tips..

Recalling firm
Beckman Coulter Inc.
Classification
Class II
Status
Terminated
Initiated
2018-01-18
Terminated
2023-09-27
Firm location
Brea, CA, United States

Product description

AutoMate 2550 "High Speed" Catalog Number ODL25255; AutoMate 1250 "Standard" Catalog Number ODL25125. Chemistry analyzer sample processing and sorting system.

Reason for recall

A misaligned pipette tip sensor poses the remote risk of cross-contamination when failing to cause a no-tip error, resulting in pipetting patient samples without pipette tips.