Recall Z-1150-2023— Siemens Healthcare Diagnostics, Inc.

Class II · Ongoing

Class II FDA medical device recall by Siemens Healthcare Diagnostics, Inc., initiated 2023-01-09. It names one FDA 510(k) clearance: K172201. Product: Atellica IM Folate 700 test kit- For in vitro diagnostic use in the quantitative determination of folate in human serum and red blood cells. Siemens Material Number (SMN): 10995573. Reason: Negative bias occurred when whole blood calibration (Atellica IM Fol or ADVIA Centaur FolateBA/FolBA) used to test serum samples..

Recalling firm
Siemens Healthcare Diagnostics, Inc.
Classification
Class II
Status
Ongoing
Initiated
2023-01-09
Firm location
East Walpole, MA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Atellica IM Folate 700 test kit- For in vitro diagnostic use in the quantitative determination of folate in human serum and red blood cells. Siemens Material Number (SMN): 10995573

Reason for recall

Negative bias occurred when whole blood calibration (Atellica IM Fol or ADVIA Centaur FolateBA/FolBA) used to test serum samples.