Recall Z-1144-2015— Spacelabs Healthcare Inc

Class II · Terminated

Class II FDA medical device recall by Spacelabs Healthcare Inc, initiated 2015-01-28, terminated 2015-12-30. It names one FDA 510(k) clearance: K112962. Product: Spacelabs Healthcare XPREZZON Bedside Monitor, Model 91393. The Spacelabs Healthcare XPREZZON Bedside Monitor passively displays data generated by Spacelabs parameter modules, Flexport interfaces, and other Spacelabs SDLG based products as waveform and numeric displays, trends and alarms. Key monitored parameters available on the model 91393, when employing the Spacelabs Command Module consist of EGG, respiration, invasive and noninvasive blood pressure, Sp02, temperature and cardiac output.. Reason: Spacelabs Healthcare XPREZZON Bedside Monitors, Model 91393, are recalled because the firm has received multiple reports of XPREZZON monitors failing to boot up or returning to factory default configuration settings following power on or reset..

Recalling firm
Spacelabs Healthcare Inc
Classification
Class II
Status
Terminated
Initiated
2015-01-28
Terminated
2015-12-30
Firm location
Snoqualmie, WA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Spacelabs Healthcare XPREZZON Bedside Monitor, Model 91393. The Spacelabs Healthcare XPREZZON Bedside Monitor passively displays data generated by Spacelabs parameter modules, Flexport interfaces, and other Spacelabs SDLG based products as waveform and numeric displays, trends and alarms. Key monitored parameters available on the model 91393, when employing the Spacelabs Command Module consist of EGG, respiration, invasive and noninvasive blood pressure, Sp02, temperature and cardiac output.

Reason for recall

Spacelabs Healthcare XPREZZON Bedside Monitors, Model 91393, are recalled because the firm has received multiple reports of XPREZZON monitors failing to boot up or returning to factory default configuration settings following power on or reset.

Data sourced from openFDA. This site is unofficial and independent of the FDA.