Recall Z-1136-2021— Ortho-Clinical Diagnostics, Inc.

Class II · Terminated

Class II FDA medical device recall by Ortho-Clinical Diagnostics, Inc., initiated 2021-01-14, terminated 2022-12-27. FDA's recall record names no specific 510(k) clearance for this event. Product: VITROS Chemistry Products VersaTips - Product Usage: used to dispense samples to MicroSlide assays and sample and reagent fluid for MicroWell assays. In addition, VersaTips are used to dispense R1 Reagent Fluid into cuvettes in combination with sample fluid for Micro Tips Assays.. Reason: A deformity affecting a specific lot of VITROS¿ Chemistry Products VersaTips. Deformed VersaTips results may be biased by up to +/-16% for both immunometric and competitive assays..

Recalling firm
Ortho-Clinical Diagnostics, Inc.
Classification
Class II
Status
Terminated
Initiated
2021-01-14
Terminated
2022-12-27
Firm location
Rochester, NY, United States

Product description

VITROS Chemistry Products VersaTips - Product Usage: used to dispense samples to MicroSlide assays and sample and reagent fluid for MicroWell assays. In addition, VersaTips are used to dispense R1 Reagent Fluid into cuvettes in combination with sample fluid for Micro Tips Assays.

Reason for recall

A deformity affecting a specific lot of VITROS¿ Chemistry Products VersaTips. Deformed VersaTips results may be biased by up to +/-16% for both immunometric and competitive assays.