Recall Z-1125-2019— Roche Diagnostics Corporation
Class II · Terminated
Class II FDA medical device recall by Roche Diagnostics Corporation, initiated 2018-12-14, terminated 2020-09-04. It names 2 FDA 510(k) clearances: K024098, K873049. Product: cobas c 111 with ISE analyzers analyzer used in the following assays: a)Creatinine plus ver.2 Catalog Number: 05401470190; b)Triglycerides Catalog Number: 04657594190 - Product Usage: Tina-quant Apolipoprotein A-1 ver.2 A lipoprotein test system is a device intended to measure lipoprotein in serum and plasma. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders and atherosclerosis Tina-quant Albumin Gen.2 Immunoturbidimetric assay for the quantitative, in vitro determination of albumin in human urine, serum, plasma, and CSF on Roche/Hitachi cobas c systems. Reason: COBAS INTEGRA¿ 400 plus and cobas c 111 analyzers Additional Extra Wash Cycle Configurations and/or Batch Mode Needed for Specific Tests Due to Carry-over.
- Recalling firm
- Roche Diagnostics Corporation
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2018-12-14
- Terminated
- 2020-09-04
- Firm location
- Indianapolis, IN, United States
510(k) clearances named in this recall
2 on FDA's record- K024098 — COBAS INTEGRA CREATININE PLUS VER.2 ASSAY
- K873049 — TRIGLYCERIDES GPO WITHOUT FREE GLYCEROL
K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
cobas c 111 with ISE analyzers analyzer used in the following assays: a)Creatinine plus ver.2 Catalog Number: 05401470190; b)Triglycerides Catalog Number: 04657594190 - Product Usage: Tina-quant Apolipoprotein A-1 ver.2 A lipoprotein test system is a device intended to measure lipoprotein in serum and plasma. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders and atherosclerosis Tina-quant Albumin Gen.2 Immunoturbidimetric assay for the quantitative, in vitro determination of albumin in human urine, serum, plasma, and CSF on Roche/Hitachi cobas c systems
Reason for recall
COBAS INTEGRA¿ 400 plus and cobas c 111 analyzers Additional Extra Wash Cycle Configurations and/or Batch Mode Needed for Specific Tests Due to Carry-over