Recall Z-1119-2026— ICU Medical, Inc.

Class II · Ongoing

Class II FDA medical device recall by ICU Medical, Inc., initiated 2025-12-15. It names one FDA 510(k) clearance: K242115. Product: ICU Medical Plum Duo Precision IV Pump, List Numbers: 400020403 (US), 400021003 (Philippines). Reason: Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at the piggyback therapy rate after the piggyback therapy is complete. In certain cases, as described below, the clinician may receive an Upper Hard Limit (UHL) or Lower Hard Limit (LHL) Violation message when programming a piggyback flush, which prevents the flush from being programmed..

Recalling firm
ICU Medical, Inc.
Classification
Class II
Status
Ongoing
Initiated
2025-12-15
Firm location
Lake Forest, IL, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

ICU Medical Plum Duo Precision IV Pump, List Numbers: 400020403 (US), 400021003 (Philippines)

Reason for recall

Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at the piggyback therapy rate after the piggyback therapy is complete. In certain cases, as described below, the clinician may receive an Upper Hard Limit (UHL) or Lower Hard Limit (LHL) Violation message when programming a piggyback flush, which prevents the flush from being programmed.

Data sourced from openFDA. This site is unofficial and independent of the FDA.