Recall Z-1119-2018— Medical Components, Inc dba MedComp

Class II · Terminated

Class II FDA medical device recall by Medical Components, Inc dba MedComp, initiated 2018-02-23, terminated 2020-01-10. It names one FDA 510(k) clearance: K022570. Product: REPAIR KIT, REF ARA101, UDI 884908083637 Product Usage: The Medcomp Repair Kit is indicated for use in replacing damaged female her connectors, clamps, or repairing extensions where there is a minimum of 4.5cm viable extension tubing.. Reason: There is the potential for connection failures related to the female two-part luer to hemodialysis blood tubing lines, etc. which may contribute to blood loss or air emboli..

Recalling firm
Medical Components, Inc dba MedComp
Classification
Class II
Status
Terminated
Initiated
2018-02-23
Terminated
2020-01-10
Firm location
Harleysville, PA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

REPAIR KIT, REF ARA101, UDI 884908083637 Product Usage: The Medcomp Repair Kit is indicated for use in replacing damaged female her connectors, clamps, or repairing extensions where there is a minimum of 4.5cm viable extension tubing.

Reason for recall

There is the potential for connection failures related to the female two-part luer to hemodialysis blood tubing lines, etc. which may contribute to blood loss or air emboli.

Data sourced from openFDA. This site is unofficial and independent of the FDA.