Recall Z-1119-2014— Greatbatch Medical
Class I · Terminated
Class I FDA medical device recall by Greatbatch Medical, initiated 2014-01-31, terminated 2016-10-31. FDA's recall record names no specific 510(k) clearance for this event. Product: Greatbatch Medical, Offset Cup Impactor. Part Numbers: P9485, T10243, T10281, T10287, T10484, T10491, T10507, T10600, T10604, T10753, T10861, T11209, T11340, T12121, T12230, T12288, T12305, T12391, T12699, T12767, T12861, T13344, T13480, T13642, T13722, T13999, T14155, T14384, T15311, T15752, T15822, T15949, T16117, T16427, T16604, T16611, T16661, T16829, T16934, T17062, T17203, T17238, T17321, T6318, T7821, T8042, T8043, T8044, T8087, T8088, T8093, T8095, T8177, T8184, T8277, T8310, T8333, T8468, T8487, T9109, T9196, T9205, T9316, T9348, T9360, T9556, T9557, T9747, T9894, T9954, T9955, T9999. The Standard Offset Cup Impactor is a reusable instrument used during acetabular replacement surgery to implant an acetabular cup. The Standard Offset Cup Impactor is provided non-sterile and must be sterilized prior to use in a surgery. The Standard Offset Cup Impactor is used to implant cups during many minimally invasive surgical approaches to Total Hip Arthroplasty ( ri-IA). EXPANDED 3-28-2014: The following Part Numbers are also affected: P10270, P6427, P6464, P6501, P7611, P7640, P7794, P7927, P8233, P8590, T10006, T10261, T10606, T10835, T11095, T11506, T11622, T12081, T12083, T12124, T12125, T13435, T13527, T14174, T15262, T15515, T16418, T16551, T16552, T16553, T17272, T5709, T6744, T7751, T9195.. Reason: Greatbatch Medical has initiated a global Field Corrective Action for the Standard Offset Cup Impactor since the product was first released in 2004. Greatbatch has identified through internal Sterility Assurance Level validation testing that a sterility assurance level of 10-6 is not achieved on the Standard Offset Cup Impactor when it is subjected to the steam sterilization cycles and associated dry times that are specified in the current Instructions For Use (IFU) provided with the device. The Standard Offset Cup Impactor is provided non-sterile and must be sterilized prior to use in a surgery. Greatbatch has not received any reports of deaths, illnesses, injuries, or other adverse effects associated with this issue..
- Recalling firm
- Greatbatch Medical
- Classification
- Class I
- Status
- Terminated
- Initiated
- 2014-01-31
- Terminated
- 2016-10-31
- Firm location
- Minneapolis, MN, United States
Product description
Greatbatch Medical, Offset Cup Impactor. Part Numbers: P9485, T10243, T10281, T10287, T10484, T10491, T10507, T10600, T10604, T10753, T10861, T11209, T11340, T12121, T12230, T12288, T12305, T12391, T12699, T12767, T12861, T13344, T13480, T13642, T13722, T13999, T14155, T14384, T15311, T15752, T15822, T15949, T16117, T16427, T16604, T16611, T16661, T16829, T16934, T17062, T17203, T17238, T17321, T6318, T7821, T8042, T8043, T8044, T8087, T8088, T8093, T8095, T8177, T8184, T8277, T8310, T8333, T8468, T8487, T9109, T9196, T9205, T9316, T9348, T9360, T9556, T9557, T9747, T9894, T9954, T9955, T9999. The Standard Offset Cup Impactor is a reusable instrument used during acetabular replacement surgery to implant an acetabular cup. The Standard Offset Cup Impactor is provided non-sterile and must be sterilized prior to use in a surgery. The Standard Offset Cup Impactor is used to implant cups during many minimally invasive surgical approaches to Total Hip Arthroplasty ( ri-IA). EXPANDED 3-28-2014: The following Part Numbers are also affected: P10270, P6427, P6464, P6501, P7611, P7640, P7794, P7927, P8233, P8590, T10006, T10261, T10606, T10835, T11095, T11506, T11622, T12081, T12083, T12124, T12125, T13435, T13527, T14174, T15262, T15515, T16418, T16551, T16552, T16553, T17272, T5709, T6744, T7751, T9195.
Reason for recall
Greatbatch Medical has initiated a global Field Corrective Action for the Standard Offset Cup Impactor since the product was first released in 2004. Greatbatch has identified through internal Sterility Assurance Level validation testing that a sterility assurance level of 10-6 is not achieved on the Standard Offset Cup Impactor when it is subjected to the steam sterilization cycles and associated dry times that are specified in the current Instructions For Use (IFU) provided with the device. The Standard Offset Cup Impactor is provided non-sterile and must be sterilized prior to use in a surgery. Greatbatch has not received any reports of deaths, illnesses, injuries, or other adverse effects associated with this issue.
Data sourced from openFDA. This site is unofficial and independent of the FDA.