Recall Z-1117-2017— Ortho-Clinical Diagnostics
Class II · Terminated
Class II FDA medical device recall by Ortho-Clinical Diagnostics, initiated 2016-12-20, terminated 2018-06-22. It names one FDA 510(k) clearance: K081411. Product: VITROS Na+ Slides, (UPN 10758750004812) For in vitro diagnostic use only. VITROS Na+ Slides quantitatively measure sodium (Na+) concentration in serum, plasma, and urine using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.. Reason: There is a potential for biased results to be generated over the 10 day on-analyzer storage limit when using VITROS Na+ Slide cartridges warmed at room temperature and loaded onto the analyzer after using the minimum warm up requirements stated in the IFU. If Quality Control (QC) results are within acceptable limits, reported results are acceptable and were not affected by this issue..
- Recalling firm
- Ortho-Clinical Diagnostics
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2016-12-20
- Terminated
- 2018-06-22
- Firm location
- Rochester, NY, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
VITROS Na+ Slides, (UPN 10758750004812) For in vitro diagnostic use only. VITROS Na+ Slides quantitatively measure sodium (Na+) concentration in serum, plasma, and urine using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.
Reason for recall
There is a potential for biased results to be generated over the 10 day on-analyzer storage limit when using VITROS Na+ Slide cartridges warmed at room temperature and loaded onto the analyzer after using the minimum warm up requirements stated in the IFU. If Quality Control (QC) results are within acceptable limits, reported results are acceptable and were not affected by this issue.