Recall Z-1108-2026— RAYSEARCH LABORATORIES AB

Class II · Ongoing

Class II FDA medical device recall by RAYSEARCH LABORATORIES AB, initiated 2025-11-28. FDA's recall record names no specific 510(k) clearance for this event. Product: Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 15.2.0, 15.1.3 Software Version: RayStation 2024A SP1, RayStation 2024A SP2 Product Description: Radiation Therapy Treatment Planning System Component: No. Reason: Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended..

Recalling firm
RAYSEARCH LABORATORIES AB
Classification
Class II
Status
Ongoing
Initiated
2025-11-28
Firm location
Stockholm, Sweden

Product description

Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 15.2.0, 15.1.3 Software Version: RayStation 2024A SP1, RayStation 2024A SP2 Product Description: Radiation Therapy Treatment Planning System Component: No

Reason for recall

Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.

Data sourced from openFDA. This site is unofficial and independent of the FDA.