Recall Z-1108-2017— PerkinElmer Life and Analytical Sciences, Wallac, OY

Class II · Terminated

Class II FDA medical device recall by PerkinElmer Life and Analytical Sciences, Wallac, OY, initiated 2017-01-05, terminated 2017-08-24. FDA's recall record names no specific 510(k) clearance for this event. Product: Spare Disk Remover Module; Product Number: 61001820 Intended for in vitro quantitative and qualitative determination of analytes in body fluids.. Reason: The firm became aware that the ball bushing axels are defective in the GSP instrument. During exposure to humid conditions, there is a risk that the axle material will corrode. Deterioration of the axle can be observed as deposits of corrosion on the defective axles. Ball bushing axles may corrode and corrosion particle may cause contamination in the wells or the corrosion will cause movement issues and possibly spillage and causing contamination of the plate wells..

Recalling firm
PerkinElmer Life and Analytical Sciences, Wallac, OY
Classification
Class II
Status
Terminated
Initiated
2017-01-05
Terminated
2017-08-24
Firm location
Turku, N/A, Finland

Product description

Spare Disk Remover Module; Product Number: 61001820 Intended for in vitro quantitative and qualitative determination of analytes in body fluids.

Reason for recall

The firm became aware that the ball bushing axels are defective in the GSP instrument. During exposure to humid conditions, there is a risk that the axle material will corrode. Deterioration of the axle can be observed as deposits of corrosion on the defective axles. Ball bushing axles may corrode and corrosion particle may cause contamination in the wells or the corrosion will cause movement issues and possibly spillage and causing contamination of the plate wells.

Data sourced from openFDA. This site is unofficial and independent of the FDA.