Recall Z-1104-2021— Draegar Medical Systems, Inc.
Class II · Ongoing
Class II FDA medical device recall by Draegar Medical Systems, Inc., initiated 2020-12-23. It names 2 FDA 510(k) clearances: K113798, K093788. Product: Infinity Acute Care System (IACS) Monitoring Solution: The IACS is intended for multi-parameter, physiologic patient monitoring of adult, pediatric and neonatal patients in environments where patient care is provided by trained healthcare professionals.. Reason: The Infinity M540 patient monitor may randomly reboot due to an error to correctly transmit and read the header data of files in the memory of the device. Under this situation, the device will try to reboot to mitigate the error. The device will be available again for use within 30 seconds. If this error continues and the M540 reboots three times in a time span of 10 minutes, it will enter a fail-state. A fail-state will annunciate itself with a continuous sound to alert the user. The M540 will reset to factory default and the user can manually configure and readmit the patient to continue patient monitoring..
- Recalling firm
- Draegar Medical Systems, Inc.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2020-12-23
- Firm location
- Andover, MA, United States
510(k) clearances named in this recall
2 on FDA's record- K113798 — INFINITY ACUTE CARE SYSTEM, INFINITY MCABLE- MASIMO RAINBOW SET
- K093788 — INFINITY ACUTE CARE SYSTEM (IACS) MONITORING SOLUTION
K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Infinity Acute Care System (IACS) Monitoring Solution: The IACS is intended for multi-parameter, physiologic patient monitoring of adult, pediatric and neonatal patients in environments where patient care is provided by trained healthcare professionals.
Reason for recall
The Infinity M540 patient monitor may randomly reboot due to an error to correctly transmit and read the header data of files in the memory of the device. Under this situation, the device will try to reboot to mitigate the error. The device will be available again for use within 30 seconds. If this error continues and the M540 reboots three times in a time span of 10 minutes, it will enter a fail-state. A fail-state will annunciate itself with a continuous sound to alert the user. The M540 will reset to factory default and the user can manually configure and readmit the patient to continue patient monitoring.