Recall Z-1101-2025— Nobel Biocare

Class II · Ongoing

Class II FDA medical device recall by Nobel Biocare, initiated 2025-01-10. It names one FDA 510(k) clearance: K211109. Product: . Nobel Biocare N1 TiUltra TCC RP 4.0x9mm . Nobel Biocare N1 TiUltra TCC RP 4.0x11mm . Nobel Biocare N1 TiUltra TCC RP 4.0x13mm Nobel Biocare N1 TiUltra TCC implants are intended for use as an endosseous dental implant in the maxilla or mandible for anchoring or supporting dental protheses to restore chewing function. Reason: Due to manufacturing issue (unintended by-product of the injection molding process) a sharp pin that may puncture the blister packaging and breaching the sterile barrier of the finished product..

Recalling firm
Nobel Biocare
Classification
Class II
Status
Ongoing
Initiated
2025-01-10
Firm location
Goteborg, Sweden

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

. Nobel Biocare N1 TiUltra TCC RP 4.0x9mm . Nobel Biocare N1 TiUltra TCC RP 4.0x11mm . Nobel Biocare N1 TiUltra TCC RP 4.0x13mm Nobel Biocare N1 TiUltra TCC implants are intended for use as an endosseous dental implant in the maxilla or mandible for anchoring or supporting dental protheses to restore chewing function

Reason for recall

Due to manufacturing issue (unintended by-product of the injection molding process) a sharp pin that may puncture the blister packaging and breaching the sterile barrier of the finished product.

Data sourced from openFDA. This site is unofficial and independent of the FDA.