Recall Z-1100-2013— GE Healthcare, LLC

Class II · Terminated

Class II FDA medical device recall by GE Healthcare, LLC, initiated 2012-04-23, terminated 2014-03-14. It names one FDA 510(k) clearance: K103327. Product: GE Healthcare, Discovery MR750w 3.0T System. Part Number: 5352293. Product Usage: It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra,dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate,blood vessels, and musculoskeletal regions of the body.. Reason: GE Healthcare has recently become aware of a potential issue due to image flip in frequency encoding direction associated with a specific workflow of your Discovery MR750w system. Image flip issue in frequency encoding direction (including S/I, A/P and R/L) can result in incorrect localization of anatomy and pathological findings..

Recalling firm
GE Healthcare, LLC
Classification
Class II
Status
Terminated
Initiated
2012-04-23
Terminated
2014-03-14
Firm location
Waukesha, WI, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

GE Healthcare, Discovery MR750w 3.0T System. Part Number: 5352293. Product Usage: It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra,dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate,blood vessels, and musculoskeletal regions of the body.

Reason for recall

GE Healthcare has recently become aware of a potential issue due to image flip in frequency encoding direction associated with a specific workflow of your Discovery MR750w system. Image flip issue in frequency encoding direction (including S/I, A/P and R/L) can result in incorrect localization of anatomy and pathological findings.