Recall Z-1096-2025— Ethicon, LLC

Class II · Ongoing

Class II FDA medical device recall by Ethicon, LLC, initiated 2024-12-20. It names one FDA 510(k) clearance: K133356. Product: PROLENE Polypropylene Suture, REF: EH7585H, 8706H, 8833H, 8963H, 8935H PROLENE Suture is indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic, and neurosurgical procedures.. Reason: Sutures may have an open seal on the primary packaging, due to a manufacturing issue on a specific machine, and if issue is undetected could lead to breach in sterility, could introduce pathogens to the patient and cause infection..

Recalling firm
Ethicon, LLC
Classification
Class II
Status
Ongoing
Initiated
2024-12-20
Firm location
San Lorenzo, PR, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

PROLENE Polypropylene Suture, REF: EH7585H, 8706H, 8833H, 8963H, 8935H PROLENE Suture is indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic, and neurosurgical procedures.

Reason for recall

Sutures may have an open seal on the primary packaging, due to a manufacturing issue on a specific machine, and if issue is undetected could lead to breach in sterility, could introduce pathogens to the patient and cause infection.

Data sourced from openFDA. This site is unofficial and independent of the FDA.