Recall Z-1094-2013— Trophy Sas

Class II · Terminated

Class II FDA medical device recall by Trophy Sas, initiated 2012-09-18, terminated 2017-05-02. FDA's recall record names no specific 510(k) clearance for this event. Product: CareStream Dental Kodak 2100 Intraoral X-ray system Product Usage: The expected usage (intended use) for all of these products is for dental radiographic examination and diagnosis of the teeth, jaw and oral structures.. Reason: There have been approximately 2% of the x-ray generators manufactured between February 2012 and November 2012 returned due to errors. These generators showed the type of defect that indicates poor contact of HT transformer power pin. The defect is a bent connector associated with the generator..

Recalling firm
Trophy Sas
Classification
Class II
Status
Terminated
Initiated
2012-09-18
Terminated
2017-05-02
Firm location
Croissy Beaubourg, France

Product description

CareStream Dental Kodak 2100 Intraoral X-ray system Product Usage: The expected usage (intended use) for all of these products is for dental radiographic examination and diagnosis of the teeth, jaw and oral structures.

Reason for recall

There have been approximately 2% of the x-ray generators manufactured between February 2012 and November 2012 returned due to errors. These generators showed the type of defect that indicates poor contact of HT transformer power pin. The defect is a bent connector associated with the generator.

Data sourced from openFDA. This site is unofficial and independent of the FDA.