Recall Z-1092-2014— Zimmer, Inc.

Class II · Terminated

Class II FDA medical device recall by Zimmer, Inc., initiated 2014-01-10, terminated 2014-09-11. It names one FDA 510(k) clearance: K041549. Product: 00434811113, REF I4348-111-13 Zimmer Trabecular Metal Shoulder Humeral Stem 48 Degrees 11 mm Stem Diameter 130 mm Stem Length Trabecular Metal Humeral Stem 48 Degree 11mm x 130mm, Sterile, QTY 1, Rx Product Usage: Zimmer Trabecular Metal TM Humeral Stem is a shoulder prosthetic replacement device that is indicated for the treatment of severe pain or significant disability in degenerative, rheumatoid, or traumatic disease of the glenohumeral joint as well as humeral fractures. The humeral stem uses. Trabecular Metal around the proximal part of the stem for biological ingrowth.. Reason: This single manufacturing lot has the potential for the bond strength between the Trabecular Metal material and substrate to be below the minimum required specification. Testing found bond strength approximately 80% of the minimum required by specification in one of the three zones of the Trabecular Metal. The other two zones exceeded the minimum specification..

Recalling firm
Zimmer, Inc.
Classification
Class II
Status
Terminated
Initiated
2014-01-10
Terminated
2014-09-11
Firm location
Warsaw, IN, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

00434811113, REF I4348-111-13 Zimmer Trabecular Metal Shoulder Humeral Stem 48 Degrees 11 mm Stem Diameter 130 mm Stem Length Trabecular Metal Humeral Stem 48 Degree 11mm x 130mm, Sterile, QTY 1, Rx Product Usage: Zimmer Trabecular Metal TM Humeral Stem is a shoulder prosthetic replacement device that is indicated for the treatment of severe pain or significant disability in degenerative, rheumatoid, or traumatic disease of the glenohumeral joint as well as humeral fractures. The humeral stem uses. Trabecular Metal around the proximal part of the stem for biological ingrowth.

Reason for recall

This single manufacturing lot has the potential for the bond strength between the Trabecular Metal material and substrate to be below the minimum required specification. Testing found bond strength approximately 80% of the minimum required by specification in one of the three zones of the Trabecular Metal. The other two zones exceeded the minimum specification.