Recall Z-1089-2026— Medline Industries, LP
Class II · Ongoing
Class II FDA medical device recall by Medline Industries, LP, initiated 2025-12-16. FDA's recall record names no specific 510(k) clearance for this event. Product: Medline medical procedure convenience kits labeled as: 1) LAVH PROCEDURE, REF CDS980754T; 2) GYN LAP HYSTERECTOMY CDS, REF CDS982662Q; 3) DAVINCI HYSTERECTOMY CDS, REF CDS985344F; 4) LAVH, REF DYNJQ9041R; 5) LAPAROSCOPIC HYSTERECTOMY, REF DYNJ900244I; 6) ROBOTIC LAVH, REF DYNJ901281I; 7) LAVH-LF, REF DYNJ901829T; 8) LEX LAP HYST, REF DYNJ902030J; 9) LITHOTOMY-SLINGS-LF, REF DYNJ905274D; 10) GYN LAPAROSCOPY, REF DYNJ905485J; 11) ROBOTIC HYSTERECTOMY, REF DYNJ906380C; 12) LAPAROSCOPIC HYSTERECTOMY, REF DYNJ910634; 13) LAVH GYN/ONC, REF DYNJ910927.. Reason: Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use..
- Recalling firm
- Medline Industries, LP
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2025-12-16
- Firm location
- Northfield, IL, United States
Product description
Medline medical procedure convenience kits labeled as: 1) LAVH PROCEDURE, REF CDS980754T; 2) GYN LAP HYSTERECTOMY CDS, REF CDS982662Q; 3) DAVINCI HYSTERECTOMY CDS, REF CDS985344F; 4) LAVH, REF DYNJQ9041R; 5) LAPAROSCOPIC HYSTERECTOMY, REF DYNJ900244I; 6) ROBOTIC LAVH, REF DYNJ901281I; 7) LAVH-LF, REF DYNJ901829T; 8) LEX LAP HYST, REF DYNJ902030J; 9) LITHOTOMY-SLINGS-LF, REF DYNJ905274D; 10) GYN LAPAROSCOPY, REF DYNJ905485J; 11) ROBOTIC HYSTERECTOMY, REF DYNJ906380C; 12) LAPAROSCOPIC HYSTERECTOMY, REF DYNJ910634; 13) LAVH GYN/ONC, REF DYNJ910927.
Reason for recall
Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.
Data sourced from openFDA. This site is unofficial and independent of the FDA.