Recall Z-1089-2015— Biomet Spine, LLC

Class II · Terminated

Class II FDA medical device recall by Biomet Spine, LLC, initiated 2011-12-19, terminated 2015-03-23. FDA's recall record names no specific 510(k) clearance for this event. Product: Telluride Percutaneous Rod Inserter Long, Model #7706-1904. The Inserter is a reusable device provided within an aluminum sterilization case along with other orthopedic manual surgical instruments. The case consists of three trays of instruments, and the middle tray contains a quantity of two Inserters. The Inserter is for use as part of a Telluride System pedicle screw procedure.. Reason: The Telluride Percutaneous Rod Inserter Long does not contain a feature that locks the Rod in place during use. There is a potential for the Rod to disengage from the Inserter during insertion, and could lead to harm to the patient..

Recalling firm
Biomet Spine, LLC
Classification
Class II
Status
Terminated
Initiated
2011-12-19
Terminated
2015-03-23
Firm location
Broomfield, CO, United States

Product description

Telluride Percutaneous Rod Inserter Long, Model #7706-1904. The Inserter is a reusable device provided within an aluminum sterilization case along with other orthopedic manual surgical instruments. The case consists of three trays of instruments, and the middle tray contains a quantity of two Inserters. The Inserter is for use as part of a Telluride System pedicle screw procedure.

Reason for recall

The Telluride Percutaneous Rod Inserter Long does not contain a feature that locks the Rod in place during use. There is a potential for the Rod to disengage from the Inserter during insertion, and could lead to harm to the patient.

Data sourced from openFDA. This site is unofficial and independent of the FDA.