Recall Z-1088-2016— Medtronic Navigation, Inc.
Class II · Terminated
Class II FDA medical device recall by Medtronic Navigation, Inc., initiated 2016-01-29, terminated 2016-03-10. It names one FDA 510(k) clearance: K151000. Product: Medtronic 0-arm 02 Imaging System Catalog Number: Bl-700-02000 The O-arm O2 Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The O-arm O2 Imaging System is compatible with certain image guided surgery systems.. Reason: Screws holding the detector panel in as a result the detector panel could make contact with other parts with in the 0-arm¿ 02 Imaging System, causing noise and vibration when being positioned or during 3D Imaging could lead to poor image quality.
- Recalling firm
- Medtronic Navigation, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2016-01-29
- Terminated
- 2016-03-10
- Firm location
- Littleton, MA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Medtronic 0-arm 02 Imaging System Catalog Number: Bl-700-02000 The O-arm O2 Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The O-arm O2 Imaging System is compatible with certain image guided surgery systems.
Reason for recall
Screws holding the detector panel in as a result the detector panel could make contact with other parts with in the 0-arm¿ 02 Imaging System, causing noise and vibration when being positioned or during 3D Imaging could lead to poor image quality