Recall Z-1087-2026— Medline Industries, LP
Class II · Ongoing
Class II FDA medical device recall by Medline Industries, LP, initiated 2025-12-16. FDA's recall record names no specific 510(k) clearance for this event. Product: Medline medical procedure convenience kits labeled as: 1) WCH BARI-GASTRECTOMY CDS, REF CDS860237AF; 2) ROBOTIC PROCEDURE, REF CDS982128Q; 3) CUH THORACIC, REF CDS984338M; 4) THORACIC PROCEDURE, REF CDS984640I; 5) DAVINCI ROBOTIC PACK-LF, REF DYNJ0525321O; 6) LAP CHOLE PACK, REF DYNJ57640F; 7) GYN/PROSTATE ROBOTIC PACK, REF DYNJ62594D; 8) RR-DAVINCI PROSTATE ACCESSORY, REF DYNJ63392D; 9) LAPAROSCOPY PACK, REF DYNJ68187B; 10) LAPAROSCOPY PACK, REF DYNJ82368A; 11) ROBOTIC KIDNEY PACK, REF DYNJ85261; 12) DOWNTOWN ROBOTICS, REF DYNJ86178B; 13) DAVINCI LITHOTOMY PACK, REF DYNJ87610; 14) ROBOTIC WHIPPLE PACK, REF DYNJ88433; 15) PROSTATE DAVINCI, REF DYNJ900127G; 16) GENERAL LAPAROSCOPY, REF DYNJ901847M; 17) LAG TLH, REF DYNJ901890N; 18) LITHOTOMY ROBOTIC CDS, REF DYNJ903948S; 19) LAP, REF DYNJ904576B; 20) ROBOTIC, REF DYNJ905147B; 21) ROBOTICS GYN, REF DYNJ905824M; 22) WH DAVINCI TLH PROCEDURE, REF DYNJ905920B; 23) CHN OR BASIC ROBOTIC, REF DYNJ905982I; 24) KIT ROBOTIC RFD, REF DYNJ906311F; 25) ROBOTIC PROSTATE, REF DYNJ906381G; 26) DAVINCI, REF DYNJ907392C; 27) CHIASSON GASTRIC SLEEVE KIT, REF DYNJ907405C; 28) GYN, REF DYNJ908504B; 29) UROLOGY ROBOTIC, REF DYNJ908588I; 30) THORACIC ROBOTICS, REF DYNJ908777A; 31) GENERAL ROBOT, REF DYNJ908969A; 32) KIT ROBOTIC, REF DYNJ909870B; 33) ROBOTIC THORACOSCOPY, REF DYNJ910924.. Reason: Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use..
- Recalling firm
- Medline Industries, LP
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2025-12-16
- Firm location
- Northfield, IL, United States
Product description
Medline medical procedure convenience kits labeled as: 1) WCH BARI-GASTRECTOMY CDS, REF CDS860237AF; 2) ROBOTIC PROCEDURE, REF CDS982128Q; 3) CUH THORACIC, REF CDS984338M; 4) THORACIC PROCEDURE, REF CDS984640I; 5) DAVINCI ROBOTIC PACK-LF, REF DYNJ0525321O; 6) LAP CHOLE PACK, REF DYNJ57640F; 7) GYN/PROSTATE ROBOTIC PACK, REF DYNJ62594D; 8) RR-DAVINCI PROSTATE ACCESSORY, REF DYNJ63392D; 9) LAPAROSCOPY PACK, REF DYNJ68187B; 10) LAPAROSCOPY PACK, REF DYNJ82368A; 11) ROBOTIC KIDNEY PACK, REF DYNJ85261; 12) DOWNTOWN ROBOTICS, REF DYNJ86178B; 13) DAVINCI LITHOTOMY PACK, REF DYNJ87610; 14) ROBOTIC WHIPPLE PACK, REF DYNJ88433; 15) PROSTATE DAVINCI, REF DYNJ900127G; 16) GENERAL LAPAROSCOPY, REF DYNJ901847M; 17) LAG TLH, REF DYNJ901890N; 18) LITHOTOMY ROBOTIC CDS, REF DYNJ903948S; 19) LAP, REF DYNJ904576B; 20) ROBOTIC, REF DYNJ905147B; 21) ROBOTICS GYN, REF DYNJ905824M; 22) WH DAVINCI TLH PROCEDURE, REF DYNJ905920B; 23) CHN OR BASIC ROBOTIC, REF DYNJ905982I; 24) KIT ROBOTIC RFD, REF DYNJ906311F; 25) ROBOTIC PROSTATE, REF DYNJ906381G; 26) DAVINCI, REF DYNJ907392C; 27) CHIASSON GASTRIC SLEEVE KIT, REF DYNJ907405C; 28) GYN, REF DYNJ908504B; 29) UROLOGY ROBOTIC, REF DYNJ908588I; 30) THORACIC ROBOTICS, REF DYNJ908777A; 31) GENERAL ROBOT, REF DYNJ908969A; 32) KIT ROBOTIC, REF DYNJ909870B; 33) ROBOTIC THORACOSCOPY, REF DYNJ910924.
Reason for recall
Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.
Data sourced from openFDA. This site is unofficial and independent of the FDA.