Recall Z-1082-2019— Brainlab AG

Class I · Terminated

Class I FDA medical device recall by Brainlab AG, initiated 2019-02-27, terminated 2019-07-10. It names one FDA 510(k) clearance: K070106. Product: Spine & Trauma 3D Navigation 1.0 (subpart of the system Navigation Software Spine & Trauma 3D, Version 3.0 ) Intended Use The Spine & Trauma Navigation System is intended as an intraoperative image-guided localization system to enable minimally invasive surgery.. Reason: In certain occurrences, the affected navigation software application might unexpectedly display a navigated instrument in an axial, coronal/sagittal (ACS) view representation with fixed planes in the image reconstruction (the not-updated ACS view ), instead of displaying the desirable view representation Inline View , which is commonly used for navigating invasive instruments at the spine. This could lead the surgeon to be unable to determine the position of the navigated instrument. This might occur after a crash restore or after changing between different navigation workflows during the same patient treatment..

Recalling firm
Brainlab AG
Classification
Class I
Status
Terminated
Initiated
2019-02-27
Terminated
2019-07-10
Firm location
Munich, N/A, Germany

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Spine & Trauma 3D Navigation 1.0 (subpart of the system Navigation Software Spine & Trauma 3D, Version 3.0 ) Intended Use The Spine & Trauma Navigation System is intended as an intraoperative image-guided localization system to enable minimally invasive surgery.

Reason for recall

In certain occurrences, the affected navigation software application might unexpectedly display a navigated instrument in an axial, coronal/sagittal (ACS) view representation with fixed planes in the image reconstruction (the not-updated ACS view ), instead of displaying the desirable view representation Inline View , which is commonly used for navigating invasive instruments at the spine. This could lead the surgeon to be unable to determine the position of the navigated instrument. This might occur after a crash restore or after changing between different navigation workflows during the same patient treatment.