Recall Z-1081-2026— Heraeus Medical GmbH (Dental Division)

Class II · Ongoing

Class II FDA medical device recall by Heraeus Medical GmbH (Dental Division), initiated 2025-12-09. It names one FDA 510(k) clearance: K210607. Product: Brand Name: PALACOS Product Name: PALACOS R+G pro 40; PALACOS R+G pro 80 Model/Catalog Number: 5081288; 5081289 Software Version: N/A Product Description: Radiopaque, poly(methyl methacrylate)-based (PMMA) bone cement, pre-filled into a mixing and application system, suitable for use with or without vacuum (ready to mix). Component: N/A. Reason: increase of complaints regarding the breakage of the ampoules within their device systems. An extensive investigation revealed that multiple factors have an impact on the ampoule breakage. An unfortunate combination of these influence factors can lead to unbreakable ampoules. If none or not all ampoules break, the product cannot be used on the patient since no or no appropriate cement dough is formed..

Recalling firm
Heraeus Medical GmbH (Dental Division)
Classification
Class II
Status
Ongoing
Initiated
2025-12-09
Firm location
Wehrheim, Germany

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Brand Name: PALACOS Product Name: PALACOS R+G pro 40; PALACOS R+G pro 80 Model/Catalog Number: 5081288; 5081289 Software Version: N/A Product Description: Radiopaque, poly(methyl methacrylate)-based (PMMA) bone cement, pre-filled into a mixing and application system, suitable for use with or without vacuum (ready to mix). Component: N/A

Reason for recall

increase of complaints regarding the breakage of the ampoules within their device systems. An extensive investigation revealed that multiple factors have an impact on the ampoule breakage. An unfortunate combination of these influence factors can lead to unbreakable ampoules. If none or not all ampoules break, the product cannot be used on the patient since no or no appropriate cement dough is formed.

Data sourced from openFDA. This site is unofficial and independent of the FDA.