Recall Z-1081-2025— PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Class II · Ongoing

Class II FDA medical device recall by PHILIPS MEDICAL SYSTEMS NEDERLAND B.V., initiated 2025-01-13. It names one FDA 510(k) clearance: K162859. Product: AlluraXper FD20 Biplane System Code: (1) 722008 (2) 722013. Reason: Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel.

Recalling firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Classification
Class II
Status
Ongoing
Initiated
2025-01-13
Firm location
Best, Netherlands

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

AlluraXper FD20 Biplane System Code: (1) 722008 (2) 722013

Reason for recall

Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel

Data sourced from openFDA. This site is unofficial and independent of the FDA.