Recall Z-1078-2019— Beckman Coulter Inc.
Class II · Ongoing
Class II FDA medical device recall by Beckman Coulter Inc., initiated 2018-12-05. It names one FDA 510(k) clearance: K042291. Product: UniCel DxC 660i, 680i, 860i, and 880i SYNCHRON Access Clinical Systems, that are re-configured so that the DxI system is separated from the DxC system and the DxI instrument is then connected to an automation line. System Name/ Catalog Number: DxC 660i/ A64871, DxC 680i /A64903, DxC 860i /A64935, DxC 880i /A59102. Reason: When racks are front-loaded onto the sample presentation unit while samples from the automation line are being processed, a scheduling error within the software can cause mishandling of samples during the creation of an aliquot and lead to incorrect results being generated. Affected samples typically yield non-reproducible falsely decreased results by up to 100%, however the potential for falsely elevated results cannot be ruled out..
- Recalling firm
- Beckman Coulter Inc.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2018-12-05
- Firm location
- Brea, CA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
UniCel DxC 660i, 680i, 860i, and 880i SYNCHRON Access Clinical Systems, that are re-configured so that the DxI system is separated from the DxC system and the DxI instrument is then connected to an automation line. System Name/ Catalog Number: DxC 660i/ A64871, DxC 680i /A64903, DxC 860i /A64935, DxC 880i /A59102
Reason for recall
When racks are front-loaded onto the sample presentation unit while samples from the automation line are being processed, a scheduling error within the software can cause mishandling of samples during the creation of an aliquot and lead to incorrect results being generated. Affected samples typically yield non-reproducible falsely decreased results by up to 100%, however the potential for falsely elevated results cannot be ruled out.
Data sourced from openFDA. This site is unofficial and independent of the FDA.