Recall Z-1077-2019— Beckman Coulter Inc.

Class II · Ongoing

Class II FDA medical device recall by Beckman Coulter Inc., initiated 2018-12-05. It names one FDA 510(k) clearance: K023764. Product: The UniCel DxI 600 and 800 Access Immunoassay Systems, that are connected to an automation line. UniCel DxI 600 Catalog Numbers: A30260, A71460, A71461, A92060. Unicel DxI 800 Catalog Numbers: 973100, A71456, A71457, A84545, A25288, A25285.. Reason: When racks are front-loaded onto the sample presentation unit while samples from the automation line are being processed, a scheduling error within the software can cause mishandling of samples during the creation of an aliquot and lead to incorrect results being generated. Affected samples typically yield non-reproducible falsely decreased results by up to 100%, however the potential for falsely elevated results cannot be ruled out..

Recalling firm
Beckman Coulter Inc.
Classification
Class II
Status
Ongoing
Initiated
2018-12-05
Firm location
Brea, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

The UniCel DxI 600 and 800 Access Immunoassay Systems, that are connected to an automation line. UniCel DxI 600 Catalog Numbers: A30260, A71460, A71461, A92060. Unicel DxI 800 Catalog Numbers: 973100, A71456, A71457, A84545, A25288, A25285.

Reason for recall

When racks are front-loaded onto the sample presentation unit while samples from the automation line are being processed, a scheduling error within the software can cause mishandling of samples during the creation of an aliquot and lead to incorrect results being generated. Affected samples typically yield non-reproducible falsely decreased results by up to 100%, however the potential for falsely elevated results cannot be ruled out.

Data sourced from openFDA. This site is unofficial and independent of the FDA.