Recall Z-1073-2024— Philips North America

Class II · Ongoing

Class II FDA medical device recall by Philips North America, initiated 2023-12-15. It names one FDA 510(k) clearance: K212837. Product: ProxiDiagnost upgrade- A Multi-functional general Radiography and Fluoroscopy (R/F) system Ref:706150. Reason: Issue 1: Mixed Images: The Radio Fluoroscopy (RF) viewer may also display a previous patients radiography (RAD) images when starting the next patient scan while the previous patient image export is still processing. If the issue occurs, there is a potential for incorrect patient data to be displayed in the image. Issue 2: Detector Access point: Security vulnerability specific to the Wireless Portable Detector configuration items in Philips Support Connect (PSC). Due to this vulnerability it is possible, with physical access to the system and knowledge of specific settings, to modify and export data to removable media (example: USB).

Recalling firm
Philips North America
Classification
Class II
Status
Ongoing
Initiated
2023-12-15
Firm location
Cambridge, MA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

ProxiDiagnost upgrade- A Multi-functional general Radiography and Fluoroscopy (R/F) system Ref:706150

Reason for recall

Issue 1: Mixed Images: The Radio Fluoroscopy (RF) viewer may also display a previous patients radiography (RAD) images when starting the next patient scan while the previous patient image export is still processing. If the issue occurs, there is a potential for incorrect patient data to be displayed in the image. Issue 2: Detector Access point: Security vulnerability specific to the Wireless Portable Detector configuration items in Philips Support Connect (PSC). Due to this vulnerability it is possible, with physical access to the system and knowledge of specific settings, to modify and export data to removable media (example: USB)

Data sourced from openFDA. This site is unofficial and independent of the FDA.