Recall Z-1073-2013— Siemens Healthcare Diagnostics, Inc.
Class II · Terminated
Class II FDA medical device recall by Siemens Healthcare Diagnostics, Inc., initiated 2013-03-11, terminated 2013-12-24. It names one FDA 510(k) clearance: K863594. Product: Dade(R) Actin(R) FSL Activated PTT Reagent. For use in the determination of the activated partial thromboplastin time (APTT).. Reason: Siemens has observed that the listed lot numbers show an increase in Heparin sensitivity over the shelf life which is demonstrated by testing with Dade(R) Citrol(R) Heparin Control, low and high..
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2013-03-11
- Terminated
- 2013-12-24
- Firm location
- Newark, DE, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Dade(R) Actin(R) FSL Activated PTT Reagent. For use in the determination of the activated partial thromboplastin time (APTT).
Reason for recall
Siemens has observed that the listed lot numbers show an increase in Heparin sensitivity over the shelf life which is demonstrated by testing with Dade(R) Citrol(R) Heparin Control, low and high.