Recall Z-1072-2024— Philips North America
Class II · Ongoing
Class II FDA medical device recall by Philips North America, initiated 2023-12-15. It names one FDA 510(k) clearance: K212837. Product: ProxiDiagnost N90- A Multi-functional general Radiography and Fluoroscopy (R/F) system Ref: (1)706100 (2)706110. Reason: Issue 1: Mixed Images: The Radio Fluoroscopy (RF) viewer may also display a previous patient s radiography (RAD) images when starting the next patient scan while the previous patient image export is still processing. If the issue occurs, there is a potential for incorrect patient data to be displayed in the image. Issue 2: Detector Access point: Security vulnerability specific to the Wireless Portable Detector configuration items in Philips Support Connect (PSC). Due to this vulnerability it is possible, with physical access to the system and knowledge of specific settings, to modify and export data to removable media (example: USB).
- Recalling firm
- Philips North America
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2023-12-15
- Firm location
- Cambridge, MA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
ProxiDiagnost N90- A Multi-functional general Radiography and Fluoroscopy (R/F) system Ref: (1)706100 (2)706110
Reason for recall
Issue 1: Mixed Images: The Radio Fluoroscopy (RF) viewer may also display a previous patient s radiography (RAD) images when starting the next patient scan while the previous patient image export is still processing. If the issue occurs, there is a potential for incorrect patient data to be displayed in the image. Issue 2: Detector Access point: Security vulnerability specific to the Wireless Portable Detector configuration items in Philips Support Connect (PSC). Due to this vulnerability it is possible, with physical access to the system and knowledge of specific settings, to modify and export data to removable media (example: USB)