Recall Z-1067-2016— Beckman Coulter Inc.

Class II · Terminated

Class II FDA medical device recall by Beckman Coulter Inc., initiated 2016-01-13, terminated 2016-08-29. It names one FDA 510(k) clearance: K101852. Product: iChem VELOCITY Automated Urine Chemistry, Catalog No. 800-3049, 800-3050, 800-3079, 800-3080, 800-3530, 800-7167, 800-7190, 800-7713, 800-7714, 800-7719, 800-7720 iChem VELOCITY Urine Chemistry Strips, Catalog No. 800-7212 & 800-7212-001. Reason: Iris International noted loose or missing analyte pads found in several locations (strip vial, strip provider module, strip conveyor system or waste container) that are undetectable by the iChemVELOCITY System..

Recalling firm
Beckman Coulter Inc.
Classification
Class II
Status
Terminated
Initiated
2016-01-13
Terminated
2016-08-29
Firm location
Brea, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

iChem VELOCITY Automated Urine Chemistry, Catalog No. 800-3049, 800-3050, 800-3079, 800-3080, 800-3530, 800-7167, 800-7190, 800-7713, 800-7714, 800-7719, 800-7720 iChem VELOCITY Urine Chemistry Strips, Catalog No. 800-7212 & 800-7212-001

Reason for recall

Iris International noted loose or missing analyte pads found in several locations (strip vial, strip provider module, strip conveyor system or waste container) that are undetectable by the iChemVELOCITY System.