Recall Z-1066-2021— Aomori Olympus Co., Ltd.

Class II · Terminated

Class II FDA medical device recall by Aomori Olympus Co., Ltd., initiated 2021-01-04, terminated 2024-03-11. It names one FDA 510(k) clearance: K933200. Product: PBD-203-0710 BILIARY STENT, model no. PBD-203-0710 -Product Usage: Used with endoscopes for endoscopic retrograde biliary drainage.. Reason: Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used..

Recalling firm
Aomori Olympus Co., Ltd.
Classification
Class II
Status
Terminated
Initiated
2021-01-04
Terminated
2024-03-11
Firm location
Kuroishi, N/A, Japan

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

PBD-203-0710 BILIARY STENT, model no. PBD-203-0710 -Product Usage: Used with endoscopes for endoscopic retrograde biliary drainage.

Reason for recall

Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.