Recall Z-1064-2018— Zimmer Biomet, Inc.

Class II · Terminated

Class II FDA medical device recall by Zimmer Biomet, Inc., initiated 2017-12-07, terminated 2020-04-17. It names one FDA 510(k) clearance: K140891. Product: 6.5mm x 125mm x 16mm Cannulated Screw, Item Number: 110009948 (UDI: (01)00887868111327 (17)270421 (10)120910) Product Usage: The 6.5/8.0mm Cannulated Screw System is part of a series of cannulated screws ranging from 4.0 to 8.0mm. Used in Slipped capital femoral epiphysis, Pediatric femoral neck fractures, Tibial plateau fractures, SI joint disruptions, Intercondylar femur fractures, Subtalar arthrodesis, and Fixation of pelvis and iliosacral joint.. Reason: One lot of 3.5mm x 24mm Cortical Locking screws is labeled as 6.5mm x 125mm x 16mm Cannulated screws..

Recalling firm
Zimmer Biomet, Inc.
Classification
Class II
Status
Terminated
Initiated
2017-12-07
Terminated
2020-04-17
Firm location
Warsaw, IN, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

6.5mm x 125mm x 16mm Cannulated Screw, Item Number: 110009948 (UDI: (01)00887868111327 (17)270421 (10)120910) Product Usage: The 6.5/8.0mm Cannulated Screw System is part of a series of cannulated screws ranging from 4.0 to 8.0mm. Used in Slipped capital femoral epiphysis, Pediatric femoral neck fractures, Tibial plateau fractures, SI joint disruptions, Intercondylar femur fractures, Subtalar arthrodesis, and Fixation of pelvis and iliosacral joint.

Reason for recall

One lot of 3.5mm x 24mm Cortical Locking screws is labeled as 6.5mm x 125mm x 16mm Cannulated screws.

Data sourced from openFDA. This site is unofficial and independent of the FDA.