Recall Z-1062-2024— Vyaire Medical
Class I · Ongoing
Class I FDA medical device recall by Vyaire Medical, initiated 2024-01-10. FDA's recall record names no specific 510(k) clearance for this event. Product: AirLife Adult Manual Resuscitator, 40" (1.0m) Oxygen Reservoir Tubing, Adult Mask, Expiratory Filter, REF 2K8005F. Reason: Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017..
- Recalling firm
- Vyaire Medical
- Classification
- Class I
- Status
- Ongoing
- Initiated
- 2024-01-10
- Firm location
- Mettawa, IL, United States
Product description
AirLife Adult Manual Resuscitator, 40" (1.0m) Oxygen Reservoir Tubing, Adult Mask, Expiratory Filter, REF 2K8005F
Reason for recall
Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
Data sourced from openFDA. This site is unofficial and independent of the FDA.