Recall Z-1062-2019— Keystone Dental Inc

Class II · Terminated

Class II FDA medical device recall by Keystone Dental Inc, initiated 2019-02-22, terminated 2020-06-19. It names one FDA 510(k) clearance: K112795. Product: Paltop Fully Guided Surgical Kit- containing the 1.25mm Hex Driver For surgical and restorative applications prosthetic devices, such as artificial teeth Part Number: 30-70021. Reason: Paltop Surgical Kits contain 1.25mm Hex Driver(s) which may have a slightly oversized hex driving feature, make it difficult to engage screws and abutments containing the corresponding screw feature.

Recalling firm
Keystone Dental Inc
Classification
Class II
Status
Terminated
Initiated
2019-02-22
Terminated
2020-06-19
Firm location
Burlington, MA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Paltop Fully Guided Surgical Kit- containing the 1.25mm Hex Driver For surgical and restorative applications prosthetic devices, such as artificial teeth Part Number: 30-70021

Reason for recall

Paltop Surgical Kits contain 1.25mm Hex Driver(s) which may have a slightly oversized hex driving feature, make it difficult to engage screws and abutments containing the corresponding screw feature

Data sourced from openFDA. This site is unofficial and independent of the FDA.