Recall Z-1059-2022— bioMerieux, Inc.

Class II · Ongoing

Class II FDA medical device recall by bioMerieux, Inc., initiated 2022-03-29. FDA's recall record names no specific 510(k) clearance for this event. Product: VITEK 2 Systems and VITEK 2 with MYLA.. Reason: Software issue where results sent to the LIS via HL7 format for antibiotic screen tests and synergy tests do not include the user-corrected or AES-corrected interpretation. This can potentially lead to incorrect final screen/synergy test results at the LIS..

Recalling firm
bioMerieux, Inc.
Classification
Class II
Status
Ongoing
Initiated
2022-03-29
Firm location
Durham, NC, United States

Product description

VITEK 2 Systems and VITEK 2 with MYLA.

Reason for recall

Software issue where results sent to the LIS via HL7 format for antibiotic screen tests and synergy tests do not include the user-corrected or AES-corrected interpretation. This can potentially lead to incorrect final screen/synergy test results at the LIS.