Recall Z-1058-2023— GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING

Class I · Ongoing

Class I FDA medical device recall by GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING, initiated 2022-12-19. It names one FDA 510(k) clearance: K173816. Product: GE Nuclear Medicine 600/800 series systems, specifically: a) 850, b) 850 / Dismount, c) 850 3/8 inch NM Gantry, d) 860 / Dismount, e) 860 3/8 inch NM Gantry, f) 860 5/8 inch NM Gantry, g) 860 BJ, h) NM 830 3/8 inch BJ, i) NM/CT 850 3/8 inch BJ, j) NM/CT 850 3/8 inch ROW, k) NM/CT 850 3/8 inch ROW_, l) NM/CT 850 CT 850 3/8 inch, m) NM/CT 850 ES 3/8 inch, n) NM/CT 850 ES 3/8 inch ROW, o) NM/CT 860 3/8 inch, p) NM/CT 860 3/8 inch BJ, q) NM/CT 860 3/8 inch ROW, r) NM/CT 860 ADVANCE, s) NM/CT 860 advance 3/8 inch NM Gantry, t) NM/CT 860 BEYOND, u) NM/CT 860 beyond 3/8 inch NM Gantry, v) NM/CT 860 China for China, w) NM/CT 860 CT 860 3/8 inch, x) NM/CT 860 ES 3/8 inch, y) NM/CT 860 ES 3/8 inch ROW, z) NM/CT 850 3/8 inch, aa) NM/CT 850 ES 3/8 inch ROW, bb) NM/CT 860 3/8 inch CH, cc) XELERIS 4 DR WORKSTATION; Emission Computed Tomography System. Reason: A mitigation may not be correctly implemented. If this is the case, the detector can fall, leading to life-threatening bodily injury..

Recalling firm
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
Classification
Class I
Status
Ongoing
Initiated
2022-12-19
Firm location
Tirat Carmel, Israel

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

GE Nuclear Medicine 600/800 series systems, specifically: a) 850, b) 850 / Dismount, c) 850 3/8 inch NM Gantry, d) 860 / Dismount, e) 860 3/8 inch NM Gantry, f) 860 5/8 inch NM Gantry, g) 860 BJ, h) NM 830 3/8 inch BJ, i) NM/CT 850 3/8 inch BJ, j) NM/CT 850 3/8 inch ROW, k) NM/CT 850 3/8 inch ROW_, l) NM/CT 850 CT 850 3/8 inch, m) NM/CT 850 ES 3/8 inch, n) NM/CT 850 ES 3/8 inch ROW, o) NM/CT 860 3/8 inch, p) NM/CT 860 3/8 inch BJ, q) NM/CT 860 3/8 inch ROW, r) NM/CT 860 ADVANCE, s) NM/CT 860 advance 3/8 inch NM Gantry, t) NM/CT 860 BEYOND, u) NM/CT 860 beyond 3/8 inch NM Gantry, v) NM/CT 860 China for China, w) NM/CT 860 CT 860 3/8 inch, x) NM/CT 860 ES 3/8 inch, y) NM/CT 860 ES 3/8 inch ROW, z) NM/CT 850 3/8 inch, aa) NM/CT 850 ES 3/8 inch ROW, bb) NM/CT 860 3/8 inch CH, cc) XELERIS 4 DR WORKSTATION; Emission Computed Tomography System

Reason for recall

A mitigation may not be correctly implemented. If this is the case, the detector can fall, leading to life-threatening bodily injury.