Recall Z-1057-2023— GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
Class I · Ongoing
Class I FDA medical device recall by GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING, initiated 2022-12-19. It names one FDA 510(k) clearance: K153402. Product: GE Nuclear Medicine 600/800 series systems, specifically: a) 870 CZT O540+ Gantry 7.25mm, b) D670 CZT Gantry 7.25mm, c) DISCOV NM/CT 870 CZT, d) Discovery NMCT 670 CZT O540+, e) NM/CT 870 CZT CYBER DEFENCE, f) NM/CT 870 CZT O540, g) NM/CT 870 CZT WSO, h) NM/CT 870 CZT, i) NM/CT 870 CZT O540+ 7.25MM; Emission Computed Tomography System. Reason: A mitigation may not be correctly implemented. If this is the case, the detector can fall, leading to life-threatening bodily injury..
- Recalling firm
- GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
- Classification
- Class I
- Status
- Ongoing
- Initiated
- 2022-12-19
- Firm location
- Tirat Carmel, Israel
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
GE Nuclear Medicine 600/800 series systems, specifically: a) 870 CZT O540+ Gantry 7.25mm, b) D670 CZT Gantry 7.25mm, c) DISCOV NM/CT 870 CZT, d) Discovery NMCT 670 CZT O540+, e) NM/CT 870 CZT CYBER DEFENCE, f) NM/CT 870 CZT O540, g) NM/CT 870 CZT WSO, h) NM/CT 870 CZT, i) NM/CT 870 CZT O540+ 7.25MM; Emission Computed Tomography System
Reason for recall
A mitigation may not be correctly implemented. If this is the case, the detector can fall, leading to life-threatening bodily injury.