Recall Z-1056-2018— Zimmer Biomet, Inc.
Class II · Terminated
Class II FDA medical device recall by Zimmer Biomet, Inc., initiated 2017-11-29, terminated 2020-04-02. It names one FDA 510(k) clearance: K060370. Product: LPS-FLEX GSF OPT SZ C-L¿ LPS-FLEX GSF OPT SZ C-R¿ LPS-FLEX GSF OPT SZ D-L¿ LPS-FLEX GSF OPT SZ D-R¿ LPS-FLEX GSF OPT SZ E-L¿ LPS-FLEX GSF OPT SZ E-R¿ LPS-FLEX GSK OPT SZ F-L¿ LPS-FLEX GSF OPT SZ F-R¿ LPS-FLEX GSF OPT SZ G-L¿ LPS-FLEX GSF OPT SZ G-R¿. Reason: The LDPE bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the LDPE bag on the implant after it is removed from the bag..
- Recalling firm
- Zimmer Biomet, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2017-11-29
- Terminated
- 2020-04-02
- Firm location
- Warsaw, IN, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
LPS-FLEX GSF OPT SZ C-L¿ LPS-FLEX GSF OPT SZ C-R¿ LPS-FLEX GSF OPT SZ D-L¿ LPS-FLEX GSF OPT SZ D-R¿ LPS-FLEX GSF OPT SZ E-L¿ LPS-FLEX GSF OPT SZ E-R¿ LPS-FLEX GSK OPT SZ F-L¿ LPS-FLEX GSF OPT SZ F-R¿ LPS-FLEX GSF OPT SZ G-L¿ LPS-FLEX GSF OPT SZ G-R¿
Reason for recall
The LDPE bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the LDPE bag on the implant after it is removed from the bag.
Data sourced from openFDA. This site is unofficial and independent of the FDA.