Recall Z-1055-2017— The Binding Site Group, Ltd.

Class II · Terminated

Class II FDA medical device recall by The Binding Site Group, Ltd., initiated 2012-10-09, terminated 2017-04-04. FDA's recall record names no specific 510(k) clearance for this event. Product: Digital RID Plate Reader and Software Product Code: AD400. Reason: If a control ring is marked after reading, the software will not flag results that are out of the specified QC range..

Recalling firm
The Binding Site Group, Ltd.
Classification
Class II
Status
Terminated
Initiated
2012-10-09
Terminated
2017-04-04
Firm location
Birmingham, United Kingdom

Product description

Digital RID Plate Reader and Software Product Code: AD400

Reason for recall

If a control ring is marked after reading, the software will not flag results that are out of the specified QC range.