Recall Z-1054-2020— Centurion Medical Products Corporation
Class II · Terminated
Class II FDA medical device recall by Centurion Medical Products Corporation, initiated 2019-12-11, terminated 2020-09-04. FDA's recall record names no specific 510(k) clearance for this event. Product: Centurion-Surgical Tray MNS10255 DELIVERY TRAY MNS10255 DELIVERY TRAY MNS10310 MEATOTOMY TRAY MNS10310 MEATOTOMY TRAY MNS10525 DELIVERY INSTRUMENT TRAY MNS10620 PACEMAKER TRAY MNS10820 ENT/EPISTAXIS KIT MNS10920 PODIATRY NAIL PACK MNS10985 CERCLAGE REMOVAL KIT MNS11020 VAGINAL DELIVERY TRAY MNS11040 OB DELIVERY PROCEDURE PACK MNS11040 OB DELIVERY PROCEDURE PACK MNS11180 CUTDOWN TRAY MNS11380 CYSTO PROCEDURE KIT MNS11415 BIRTHING SET MNS11430 MYELOGRAM TRAY MNS11485 DERM ROOM TRAY MNS11490 VASECTOMY TRAY MNS11490 VASECTOMY TRAY MNS11605 STERILE FLUORO SSP TRAY MNS11720 OCULOPLASTICS INSTRUMENT SET MNS11780 MOHS INSTRUMENT PACK MNS11885 OB BASIC DELIVERY INSTRUMENT TRAY MNS11980 DERMATOLOGY TRAY MNS12045 LINE REPAIR KIT MNS12090 TOENAIL REMOVAL PACK MNS12140 HAND TRAY MNS12175 PACEMAKER TRAY MNS1320 E.N.T. TRAY MNS1445 VASECTOMY MNS1485 NOSEBLEED TRAY MNS1500 JOINT ASPIRATION KIT MNS1500 JOINT ASPIRATION KIT MNS1545 LATEX FREE NASAL HEMORRHAGE SET MNS1555 EPISTAXIS TRAY MNS1900 NOSEBLEED TRAY MNS1990 NASAL KIT MNS2105 EPISTAXIS TRAY MNS2170 NASAL PACKING TRAY MNS2225B MYELOGRAM TRAY MNS2225B MYELOGRAM TRAY MNS2845 U/S, BASIC BIOPSY TRAY MNS2975 ENT EPISTAXIS TRAY MNS3345A PEDIATRIC CUTDOWN TRAY MNS3450 IUD TRAY MNS3490 CHEMO KIT MNS3925 OB INSTRUMENT TRAY MNS4090A MINOR CUTDOWN TRAY MNS4110 THORACOTOMY TRAY MNS4120 NOSEBLEED TRAY MNS4215 OR VAGINAL DELIVERY TRAY MNS4360 BREAST BIOPSY KIT MNS4405 BIOPSY DERMATOLOGY KIT MNS4545 NASAL TRAY MNS4855 WOUND EXPLORATION/DPL TRAY MNS5170 NEONATAL PROCEDURE TRAY MNS5225 RENAL BIOPSY TRAY MNS5272 OB DELIVERY INSTRUMENT TRAY MNS5410 VAG DELIVERY INSTRUMENTS MNS5410 VAG DELIVERY INSTRUMENTS MNS5425 VAGINAL DELIVERY TRAY MNS5511 ARTHROGRAM TRAY MNS5575 OB DELIVERY TRAY MNS5750A VAGINAL DELIVERY SET MNS6025 GU MEATOTOMY MNS6025 GU MEATOTOMY MNS6260 BIOPSY TRAY MNS6285 MINOR PLASTIC SET MNS6395 VAGINAL DELIVERY KIT MNS6405 MULTI-PURPOSE RADIOLOGY TRAY MNS6820 RADIOLOGIST SPINAL TRAY MNS6880 BIOPSY TRAY MNS6985 PERCUTANEOUS CATHETER PLACEMENT TRAY MNS7155 TONGUE TIE KIT MNS7285 BIOPSY KIT- SHAVE MNS7285 BIOPSY KIT- SHAVE MNS7285 BIOPSY KIT- SHAVE MNS7295 BIOPSY KIT - SHAVE/PUNCH COMBO MNS7325 CATHETER REMOVAL TRAY MNS7415 NASAL HEMORRHAGE TRAY MNS7415 NASAL HEMORRHAGE TRAY MNS7575 MINOR TISSUE TRAY MNS7605 BREAST BIOPSY KIT MNS7815 DERMAL PACK MNS7900 C.T. BIOPSY TRAY MNS8010 EPISTAXIS TRAY MNS8445 SPINAL TRAY MNS8445 SPINAL TRAY MNS8475 ED TRACHEOTOMY TRAY MNS8645 BIOPSY KIT MNS8645 BIOPSY KIT MNS8915 BLOOD CULTURE COLLECTION KIT ; CVAD MNS9005 CENTRAL LINE REMOVAL TRAY MNS9106 ULTRASOUND BIOPSY TRAY MNS9106 ULTRASOUND BIOPSY TRAY MNS9170 CUTDOWN TRAY MNS9245 PACEMAKER TRAY MNS9335 IUD TRAY MNS9455 VAGINAL DELIVERY TRAY W/INSTRUMENTS MNS9480 VAGINAL DELIVERY TRAY MNS9480 VAGINAL DELIVERY TRAY MNS9625 TRACHEOSTOMY KIT MNS965 AMNIOCENTESIS TRAY MNS9710 IUD TRAY MNS9765 PORT INSERTION TRAY MNS9780 ARTHROGRAM TRAY MNS9820 OB PACK MNS9835 VAG DELIVERY TRAY MNS9835 VAG DELIVERY TRAY MNS9855 PACEMAKER INSTRUMENT TRAY MNS9870 DR. NELSON MOHS PACK MNS9870 DR. NELSON MOHS PACK MNS9935 CATH LAB INSTRUMENT KIT. Reason: Incomplete seals on the sterile package may compromise the sterility.
- Recalling firm
- Centurion Medical Products Corporation
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2019-12-11
- Terminated
- 2020-09-04
- Firm location
- Williamston, MI, United States
Product description
Centurion-Surgical Tray MNS10255 DELIVERY TRAY MNS10255 DELIVERY TRAY MNS10310 MEATOTOMY TRAY MNS10310 MEATOTOMY TRAY MNS10525 DELIVERY INSTRUMENT TRAY MNS10620 PACEMAKER TRAY MNS10820 ENT/EPISTAXIS KIT MNS10920 PODIATRY NAIL PACK MNS10985 CERCLAGE REMOVAL KIT MNS11020 VAGINAL DELIVERY TRAY MNS11040 OB DELIVERY PROCEDURE PACK MNS11040 OB DELIVERY PROCEDURE PACK MNS11180 CUTDOWN TRAY MNS11380 CYSTO PROCEDURE KIT MNS11415 BIRTHING SET MNS11430 MYELOGRAM TRAY MNS11485 DERM ROOM TRAY MNS11490 VASECTOMY TRAY MNS11490 VASECTOMY TRAY MNS11605 STERILE FLUORO SSP TRAY MNS11720 OCULOPLASTICS INSTRUMENT SET MNS11780 MOHS INSTRUMENT PACK MNS11885 OB BASIC DELIVERY INSTRUMENT TRAY MNS11980 DERMATOLOGY TRAY MNS12045 LINE REPAIR KIT MNS12090 TOENAIL REMOVAL PACK MNS12140 HAND TRAY MNS12175 PACEMAKER TRAY MNS1320 E.N.T. TRAY MNS1445 VASECTOMY MNS1485 NOSEBLEED TRAY MNS1500 JOINT ASPIRATION KIT MNS1500 JOINT ASPIRATION KIT MNS1545 LATEX FREE NASAL HEMORRHAGE SET MNS1555 EPISTAXIS TRAY MNS1900 NOSEBLEED TRAY MNS1990 NASAL KIT MNS2105 EPISTAXIS TRAY MNS2170 NASAL PACKING TRAY MNS2225B MYELOGRAM TRAY MNS2225B MYELOGRAM TRAY MNS2845 U/S, BASIC BIOPSY TRAY MNS2975 ENT EPISTAXIS TRAY MNS3345A PEDIATRIC CUTDOWN TRAY MNS3450 IUD TRAY MNS3490 CHEMO KIT MNS3925 OB INSTRUMENT TRAY MNS4090A MINOR CUTDOWN TRAY MNS4110 THORACOTOMY TRAY MNS4120 NOSEBLEED TRAY MNS4215 OR VAGINAL DELIVERY TRAY MNS4360 BREAST BIOPSY KIT MNS4405 BIOPSY DERMATOLOGY KIT MNS4545 NASAL TRAY MNS4855 WOUND EXPLORATION/DPL TRAY MNS5170 NEONATAL PROCEDURE TRAY MNS5225 RENAL BIOPSY TRAY MNS5272 OB DELIVERY INSTRUMENT TRAY MNS5410 VAG DELIVERY INSTRUMENTS MNS5410 VAG DELIVERY INSTRUMENTS MNS5425 VAGINAL DELIVERY TRAY MNS5511 ARTHROGRAM TRAY MNS5575 OB DELIVERY TRAY MNS5750A VAGINAL DELIVERY SET MNS6025 GU MEATOTOMY MNS6025 GU MEATOTOMY MNS6260 BIOPSY TRAY MNS6285 MINOR PLASTIC SET MNS6395 VAGINAL DELIVERY KIT MNS6405 MULTI-PURPOSE RADIOLOGY TRAY MNS6820 RADIOLOGIST SPINAL TRAY MNS6880 BIOPSY TRAY MNS6985 PERCUTANEOUS CATHETER PLACEMENT TRAY MNS7155 TONGUE TIE KIT MNS7285 BIOPSY KIT- SHAVE MNS7285 BIOPSY KIT- SHAVE MNS7285 BIOPSY KIT- SHAVE MNS7295 BIOPSY KIT - SHAVE/PUNCH COMBO MNS7325 CATHETER REMOVAL TRAY MNS7415 NASAL HEMORRHAGE TRAY MNS7415 NASAL HEMORRHAGE TRAY MNS7575 MINOR TISSUE TRAY MNS7605 BREAST BIOPSY KIT MNS7815 DERMAL PACK MNS7900 C.T. BIOPSY TRAY MNS8010 EPISTAXIS TRAY MNS8445 SPINAL TRAY MNS8445 SPINAL TRAY MNS8475 ED TRACHEOTOMY TRAY MNS8645 BIOPSY KIT MNS8645 BIOPSY KIT MNS8915 BLOOD CULTURE COLLECTION KIT ; CVAD MNS9005 CENTRAL LINE REMOVAL TRAY MNS9106 ULTRASOUND BIOPSY TRAY MNS9106 ULTRASOUND BIOPSY TRAY MNS9170 CUTDOWN TRAY MNS9245 PACEMAKER TRAY MNS9335 IUD TRAY MNS9455 VAGINAL DELIVERY TRAY W/INSTRUMENTS MNS9480 VAGINAL DELIVERY TRAY MNS9480 VAGINAL DELIVERY TRAY MNS9625 TRACHEOSTOMY KIT MNS965 AMNIOCENTESIS TRAY MNS9710 IUD TRAY MNS9765 PORT INSERTION TRAY MNS9780 ARTHROGRAM TRAY MNS9820 OB PACK MNS9835 VAG DELIVERY TRAY MNS9835 VAG DELIVERY TRAY MNS9855 PACEMAKER INSTRUMENT TRAY MNS9870 DR. NELSON MOHS PACK MNS9870 DR. NELSON MOHS PACK MNS9935 CATH LAB INSTRUMENT KIT
Reason for recall
Incomplete seals on the sterile package may compromise the sterility
Data sourced from openFDA. This site is unofficial and independent of the FDA.