Recall Z-1051-2015— Teleflex Medical

Class I · Terminated

Class I FDA medical device recall by Teleflex Medical, initiated 2015-01-12, terminated 2016-05-10. FDA's recall record names no specific 510(k) clearance for this event. Product: MAQUET, SERVO Humidifier 163, Rx Only, Inmed Mfg. Sdn Bhd. Catalog No: XKC01-06-8125-8, Model No. 01-06-8125-8. A heat and moisture condenser (artificial nose) is a device intended to be positioned over a tracheotomy (a surgically created opening in the throat) or tracheal tube (a tube inserted into the trachea) to warm and humidify gases breathed in by a patient.. Reason: Defects: Cracks that may occur in the center of the corrugated flex tube of the 15 mm and 22 mm ISO female connector..

Recalling firm
Teleflex Medical
Classification
Class I
Status
Terminated
Initiated
2015-01-12
Terminated
2016-05-10
Firm location
Durham, NC, United States

Product description

MAQUET, SERVO Humidifier 163, Rx Only, Inmed Mfg. Sdn Bhd. Catalog No: XKC01-06-8125-8, Model No. 01-06-8125-8. A heat and moisture condenser (artificial nose) is a device intended to be positioned over a tracheotomy (a surgically created opening in the throat) or tracheal tube (a tube inserted into the trachea) to warm and humidify gases breathed in by a patient.

Reason for recall

Defects: Cracks that may occur in the center of the corrugated flex tube of the 15 mm and 22 mm ISO female connector.