Recall Z-1050-2019— LW Scientific, Inc.
Class II · Terminated
Class II FDA medical device recall by LW Scientific, Inc., initiated 2019-02-06, terminated 2021-05-07. FDA's recall record names no specific 510(k) clearance for this event. Product: LW Scientific ZIP IQ Combo Centrifuge, ZIC-CMBO-4RT3 Rev B Product Usage: used as a general purpose centrifuge to spin down blood in hematocrit tubes with additional rotors to spin other fluids in test tubes (15ml and 50ml), microtubes, and PCR strips. Reason: The clear, plastic, polycarbonate hematocrit rotor may break apart during use and be ejected through the plastic housing, posing a safety hazard to the technician and potentially damaging the surrounding area..
- Recalling firm
- LW Scientific, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2019-02-06
- Terminated
- 2021-05-07
- Firm location
- Lawrenceville, GA, United States
Product description
LW Scientific ZIP IQ Combo Centrifuge, ZIC-CMBO-4RT3 Rev B Product Usage: used as a general purpose centrifuge to spin down blood in hematocrit tubes with additional rotors to spin other fluids in test tubes (15ml and 50ml), microtubes, and PCR strips
Reason for recall
The clear, plastic, polycarbonate hematocrit rotor may break apart during use and be ejected through the plastic housing, posing a safety hazard to the technician and potentially damaging the surrounding area.