Recall Z-1048-2021— Aomori Olympus Co., Ltd.
Class II · Terminated
Class II FDA medical device recall by Aomori Olympus Co., Ltd., initiated 2021-01-04, terminated 2024-03-11. It names one FDA 510(k) clearance: K153625. Product: NEEDLEMASTER 6MMx 25G LOWER, model no. NM-610U-0625 - Product Usage: Used with endoscopes to perform endoscopic vascular or submucosal injection in the GI tract.. Reason: Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used..
- Recalling firm
- Aomori Olympus Co., Ltd.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2021-01-04
- Terminated
- 2024-03-11
- Firm location
- Kuroishi, N/A, Japan
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
NEEDLEMASTER 6MMx 25G LOWER, model no. NM-610U-0625 - Product Usage: Used with endoscopes to perform endoscopic vascular or submucosal injection in the GI tract.
Reason for recall
Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Data sourced from openFDA. This site is unofficial and independent of the FDA.